Microscopic gel–liquid interfaces supported by hollow microneedle array for voltammetric drug detection

Authors: 

Patricia Vazquez, Grégoire Herzog, Conor O’Mahony, Joseph O’Brien, James Scully, Alan Blake, Cian O’Mathuna, Paul Galvin

Publication Type: 
Refereed Original Article
Abstract: 
This report describes a method for integration of a gel–liquid interface in hollow microneedles compatible with minimally invasive, electrochemical detection of drugs in vivo. The electrochemical sensor was characterised using cyclic voltammetry with tetraethyl ammonium. The experimental work demonstrated the detection of propranolol as a representative drug in physiological buffer with the microneedle system. A calibration curve for propranolol was built from measurements with differential pulse stripping voltammetry, indicating a sensitivity of 43 nA μM−1, a limit of detection of 50 nM and a linear range between 50 and 200 nM.
Digital Object Identifer (DOI): 
10.1016/j.snb.2014.04.080
Publication Status: 
Published
Date Accepted for Publication: 
Wednesday, 23 April, 2014
Publication Date: 
02/05/2014
Journal: 
Sensors and Actuators B: Chemical
Research Group: 
Institution: 
Tyndall National Institute
Open access repository: 
No